Experimental steps of low-density polyethylene infusion bottle puncture force tester
How to detect the puncture force of low-density polyethylene infusion bottles?
(1) Prepare several low-density polyethylene infusion bottles according to standard requirements.
(2) Clamp the sample onto the puncture test fixture and secure the fixture onto the testing equipment.
(3) Set test parameters such as test speed and sample thickness.
(4) Click on the start experiment option, and the experiment will begin. Start the instrument and puncture the puncture site on the infusion bottle at a speed of 200 m/min ± 50mm/min
(5) The equipment automatically calculates and displays the final test results.
(6) The puncture force of a plastic puncture device shall not exceed 100N, and the puncture force of a metal puncture device shall not exceed 80N.

The puncture force of low-density polyethylene infusion bottles is a routine physical indicator specified in YBB00012002-2015 "Low Density Polyethylene Infusion Bottles", and how to detect this indicator has become a focus of attention for pharmaceutical companies.

Low density polyethylene infusion bottle puncture force tester
To know how to detect the puncture force of low-density polyethylene infusion bottles, it is necessary to understand the experimental equipment used to detect this indicator, including a professional computer software operating system that can save, analyze, and print data, a large LCD display for the testing process, a professional sensor system to ensure the accuracy of the test results, a screw drive system, and an infinite speed regulation of 1-500m/min; Rapid printing of experimental results using a micro printer; Simultaneously test both puncture force and pull-out force values; The main unit of the puncture instrument, micro printer, puncture force fixture, testing software, and communication cable areLow density polyethylene infusion bottle puncture force testerBasic configuration.
According to the YBB00012002-2015 standard for low-density polyethylene infusion bottles, several samples of this product should be taken and punctured at a speed of 200 m/min ± 50mm/min using a puncture device that complies with Figures 1 and 2. The puncture force of the plastic puncture device should not exceed 100N, and the puncture force of the metal puncture device should not exceed 80N.