E-mail
hnbnyq@126.com
-
Phone
18838155385
-
Address
No. 58 Mengjin Road, Shangjie District, Zhengzhou City
Henan Bainian Instrument Co., Ltd
hnbnyq@126.com
18838155385
No. 58 Mengjin Road, Shangjie District, Zhengzhou City
Our company's new BN-ZK series intelligent digital display constant temperature water bath pot
Recently, our company's development team has developed and optimized a new product, the BN-ZK series intelligent digital display constant temperature water bath. This upgrade will enhance the product's appearance, operation panel, and functionality. This product is widely used and suitable for operations in modern petroleum, chemical, pharmaceutical and health, environmental protection, biochemistry, experimental analysis, education and research institutions.



Product Features
1、 High brightness digital display, clear parameter settings, beautiful and elegant appearance;
2、 Controlled by microprocessor, with good temperature control linearity and small fluctuations;
3、 Adopting external heating instead of traditional built-in heating tube heating, with built-in heating and over temperature sensor, automatic power-off for dry burning, safe and reliable.
4. External heating of liquid temperature is more uniform, safer, and more reliable, and can also avoid situations like long-term use of built-in heating tubes, where scale forms on the surface of the heating tube and the heating tube melts and leaks electricity.
5. The shell and insulation board are both high-strengthPC (Polycarbon) material is made by one-time injection molding, which is corrosion resistant and anti-aging, and the liner is made of 304 stainless steel by stamping.
type number |
BN-ZK2 |
BN-ZK4 |
BN-ZK6 |
Temperature control range |
room temperature+5~100℃ |
room temperature+5~100℃ |
room temperature+5~100℃ |
temperature accuracy |
±0.5℃ |
±0.5℃ |
±0.5℃ |
timing range |
0-999minute |
0-999minute |
0-999minute |
number of holes |
2 |
4 |
6 |
volume |
6.7L |
14L |
22L |
Working size(cm) |
30*15*12 |
32*30*12 |
50*30*12 |
input power |
220V50HZ |
220V50HZ |
220V50HZ |
power(W) |
450W |
800W |
1200W |
Overall dimensions(cm) |
33*240*20 |
39*36*20 |
54*40*20 |
magnetic stirring |
none |
none |
none |
Instrument operationInstructions for Use
1Power on the temperature controller and display the temperature control area on the window,2After seconds, it will enter the normal display state.
2Temperature setting: PressSETPress the key to enter the temperature setting state, and a prompt will appear on the upper row of the display window“SP”The lower row displays the temperature setting value, which can be changed to the desired temperature value by pressing the increase and decrease keys, and then pressingSETPress the key to exit this setting state, and the modified setting values will be automatically saved. At this point, the instrument is working normally.
3Normal operation does not require changing internal parameters. If you need to change internal parameters, pressSETkey3After a few seconds, enter the internal parameter area and press each buttonSETKey change parameter indicator code.
code |
name |
Instructions |
Set range |
P |
Proportional band |
if it is0It is positional control |
0~99 |
C |
Temperature measurement bias value |
Sensor error correction |
-9.9~9.9 |
F |
heating cycle |
PIDThe heating cycle |
10~99second |
E |
heating power |
Adjustable power of heater |
50~100 |
4Precautions
(1)When the instrument is working, someone needs to be on duty, and the security check and calibration cycle of this instrument is six months.
(2)Please do not work in areas with flammable, explosive, or corrosive gases.
(3)instrument displayHHHTo measure values that exceed the limit on the instrumentShort circuit in the process or sensor, instrument displayLLLThe sensor is not installed or the sensor is disconnected.
(4)There is a water level indicator inside the tank, and only an external temperature sensor is used for this water bath. Please make sure that the water level exceeds the water level indicator line inside the tank. If the liquid level in the tank is low, it will cause abnormal temperature fluctuations in the liquid inside the tank.
During normal operation, the upper row displays the measured temperature value and the lower row displays the set temperature value. When entering the internal parameters, the upper row displays the parameter code and the lower row displays the parameters.