- Phone
-
Address
No. 2466 Science and Technology Avenue, Lin'an District, Hangzhou City, Zhejiang Province, China
Hangzhou Puyu Technology Development Co., Ltd
No. 2466 Science and Technology Avenue, Lin'an District, Hangzhou City, Zhejiang Province, China

Benzoic acid is mainly used in the production of antifungal and disinfectant preservatives, fragrances, and food preservatives. In theory, benzoic acid and its sodium salts have low toxicity and can be excreted through normal metabolism in the human body at reasonable doses. Sorbic acid is another widely used preservative in food with good safety, but excessive use can affect bone development and increase liver and kidney burden.
Saccharin sodium, with a sweetness 200~700 times that of sucrose, is a commonly used artificial sweetener. It is not absorbed in the human body, but excessive intake can reduce the absorption capacity of the small intestine and decrease appetite.
In food production, some enterprises are prone to using the above-mentioned food additives beyond the scope in order to improve product taste and extend shelf life, resulting in substandard products and threatening consumer health. In order to strictly control food quality and safety, this article establishes an ultra-high performance liquid chromatography detection method for benzoic acid, sorbic acid, and sodium saccharin in food based on the standard "GB 5009.28-2016 Determination of Benzoic Acid, Sorbic Acid, and Sodium Saccharin in Food", which can provide reference for the detection of preservatives and sweeteners in food.
Pu Yu Technology Solution

01 Quick and precise control, stable and efficient
Easy to use and reliable, with a pressure resistance of up to 20000 psi, improving separation and reducing operating time.
02 Intelligent and convenient, with extremely low residue
Multi mode, wide range, low residue, ensuring stable and reliable analysis.
03 Excellent light efficiency, sensitive perception
Sub microliter waveguide flow cell, high light efficiency, high signal-to-noise ratio, meets diverse application needs.
04 Smart Interconnection, Security and Compliance
Support standalone/network deployment, local/cloud storage, ensuring data security and compliance.
Application Showcase
01 Experimental Instruments
EXPEC 5180 Ultra High Performance Liquid Chromatography (equipped with a binary ultra-high pressure infusion pump ULC 580, automatic sampler AS 580, column temperature box CH 582, diode array detector DAD 580, and solvent disk ET 580)
02 Pre processing methods
According to the GB 5009.28-2016 National Food Safety Standard for the Determination of Benzoic Acid, Sorbic Acid, and Sodium Saccharin in Foods, the sample was extracted with water, the high-fat sample was defatted with n-hexane, and the high protein sample was precipitated with protein precipitant. After centrifugation, an appropriate amount of supernatant was taken and passed through a 0.22 μ m filter membrane for liquid chromatography determination.
03 Instrument conditions

Analysis results
01 Chromatogram
The chromatograms of benzoic acid, sorbic acid, and sodium saccharin obtained using EXPEC 5180 detection are as follows:

02 Standard Curve
Dilute the standard stock solution with water step by step to obtain a series of concentrations of benzoic acid, sorbic acid, and sodium saccharin mixed standard working solutions (0.2 μ g/mL, 0.5 μ g/mL, 1 μ g/mL, 2 μ g/mL, 5 μ g/mL, 10 μ g/mL, and 20 μ g/mL). Inject samples from low to high concentration for analysis. Use the mass concentration of the target substance in the standard series solution as the horizontal axis and its corresponding peak area as the vertical axis to establish a calibration curve, with linear correlation coefficients ≥ 0.9999.

03 Method detection limit
Configure 0.5 μ g/mL standard solution for injection, calculate the quantification limit and detection limit according to S/N=10 and S/N=3:

04 Precision
Prepare mixed standard solutions of benzoic acid, sorbic acid, and sodium saccharin with concentrations of 0.5 μ g/mL, 2 μ g/mL, and 10 μ g/mL, and inject them continuously for 6 injections. Analyze the RSD of retention time and peak area of each group of samples to evaluate instrument precision. The results show that the RSD range of retention time is 0.08~0.37%, and the RSD range of peak area is 0.18~4.59%.

05 Recovery rate (accuracy)
By conducting two concentration spiked recovery experiments on blank samples, the results showed that the spiked recovery rate (n=6) ranged from 89.1% to 101.7%.

Conclusion
This article establishes a rapid and sensitive detection method for benzoic acid, sorbic acid, and sodium saccharin in food based on the EXPEC 5180 UHPLC technology. The linearity, sensitivity, precision, and accuracy of the method were examined sequentially, and the results showed that the linearity was good within the concentration range of 0.2-20 μ g/mL, with linear correlation coefficients all greater than 0.9999. Take standard solutions with concentrations of 0.5 μ g/mL, 2 μ g/mL, and 10 μ g/mL and inject them continuously for 6 times. The retention time RSD range is 0.08~0.37%, and the peak area RSD range is 0.18~4.59%, indicating good precision. In addition, two concentration levels of spiked recovery experiments were conducted on blank samples, and the spiked recovery rate (n=6) ranged from 89.1% to 101.7%, with good accuracy. The detection limit and quantification limit calculated based on the signal-to-noise ratio using standard curve low point injection are significantly better than the standard requirements.
Puyu Technology, supported by high-performance domestic analytical instruments, continues to provide efficient and reliable solutions for food additive supervision and enterprise quality control, fully safeguarding the safety of people's tongues.