No. 172, Lindong Road, Nanyuan Street, Yuhang District, Hangzhou City
The artificial climate chamber is a high-precision constant temperature and humidity light simulation test equipment with cooling, heating, and illumination control. It can be used for plant cultivation, breeding experiments, cultivation and preservation of bacteria, molds, and microorganisms, BOD determination for water analysis, and constant temperature tests for other purposes. It is an ideal experimental equipment for research institutions such as biological genetic engineering, medical, health and epidemic prevention, drug testing, agriculture, animal husbandry, and aquatic products.
Working principle: The artificial climate chamber is composed of sensors located inside the chamber, where the actual temperature and humidity sensed by the temperature and humidity sensors are converted into electrical signals. At the same time, the preset light intensity and electrical signal can be controlled by a microcomputer to operate the heater, humidifier, light intensity or refrigeration compressor, thereby achieving the required temperature, humidity and light intensity control.
Daily usage precautions and maintenance:
a) The use of artificial climate chambers requires reliable grounding. Incorrect grounding can cause electric shock or fire.
b) It should be ensured that the power plug is firmly inserted into the power board. If the plug is not tight and easy to loosen, it may also cause electric shock, overheating, or equipment failure.
c) Do not use it during the process. Unplug the power plug without turning off the device power switch to avoid fire caused by overheating or sparks.
d) Try to avoid damaging the power cord or using non specified power cords. To avoid equipment damage, electric shock, and fire hazards.
e) Do not disconnect or use extension cords in the middle of the power cord. To avoid electric shock or fire caused by overheating.
f) Do not unplug or plug the power cord with wet hands. Otherwise, there is a risk of electric shock.
g) Do not frequently change the set value during use to avoid overloading caused by frequent starting of the compressor, which may affect the service life of the equipment.
h) It is strictly prohibited to use tap water or other unclean water sources when injecting water into the water tank. Pure water, distilled water, or other ultra clean water needs to be injected. Otherwise, it may cause the humidification to malfunction and result in equipment failure.
i) Be sure to close the inner glass door before closing the box door. If the inner door is not closed properly, even if the box door is closed, the equipment may not be able to operate at high performance. Please be careful when closing the inner and outer doors, and do not use too much force to damage the door seal.
j) To ensure uniform temperature inside the box, it is necessary to regularly check whether the circulating fan inside the box is operating normally. During the experiment, items inside the box should not be placed too densely and should not block the fan outlet to facilitate airflow circulation inside the box.
k) Do not touch, collision with temperature and humidity sensors inside the box may cause temperature or humidity to lose control.
l) Do not insert fingers, sticks, or other foreign objects into the air supply or inlet. To avoid triggering the high-speed running fan and causing equipment damage or human injury.
m) If there is an abnormal state (such as burning smell), the power should be immediately turned off to terminate the operation of the equipment, or the power should be disconnected with a circuit breaker. If the abnormal state persists, it will cause overheating, fire or electric shock, and even damage the machine.
n) Do not carry out repairs without authorization. Incorrect repair can lead to fire or electric shock.