What is the method for installing and debugging a fully automatic integrated distillation apparatus?
The fully automatic integrated distillation apparatus mainly consists of four parts: heating device, distillation device, circulating cooling water device, and receiving device. The heating device is equipped with intelligent control function for heating rate, which can achieve precise control of heating temperature and heating efficiency; The circulating cooling device is equipped with cooling temperature display and control functions to ensure the cooling effect.
Distillation is a common and important pre-treatment step in water quality testing and food inspection for this product. Traditional distillation equipment has independent heating, distillation, condensation, and receiving parts, which are cumbersome to operate and often lead to distillation failure due to the lack of distillation endpoint control. The emergence of this product solves these problems perfectly.
The installation and debugging method of the fully automatic integrated distillation apparatus is as follows:
1. Check the cooling water circulation system
Turn on the cold water circulation switch, and water begins to enter the condenser tube. Observe whether there is water leakage at each interface. If there is water leakage, reseal the joint. After the condenser pipeline is filled with circulating water, the normal liquid level in the water tank is between the yellow and red lines. The normal operation of circulating water can be determined by whether the green water flow meter on the right side of the instrument is rotating.
2. Specific operation method
Distilled water is added to the distillation bottle, and the upper part is sealed and connected to the anti backflow device through a hollow sealing plug and hose. The long arm mouth is sealed and connected to the condenser bottle, and the distillate from the condenser bottle is sealed and connected to the receiving bottle through a hose.
1) Select Distillation Experiment
Turn on the main switch and auxiliary switch, enter the "preset menu" interface, press the "select" key to select "1". Press the "Confirm" button to enter the "Experiment Selection" interface, press the "Select" button to select, press the "Confirm" button to select, and press the "Return" button to return to the "Preset Menu" interface. According to the distillation endpoint control of each experiment, there are default values for weight and time settings, which users can adjust according to their own needs.
2) Select the overall power of the machine
In the "Preset Menu" interface, select "Machine Power" and press the "Confirm" button to enter. Select the desired power supply, press the "confirm" button, and return to the "preset menu" interface. This project is selected based on user needs and power supply voltage capacity. The default setting before leaving the factory is 3600W.
3. Select control mode
In the "Preset Menu" interface, select "3". Control mode ", press the" confirm "button to enter, select the desired control mode, and press the" return "button to enter the" preset menu "interface. This can be set according to user needs, and the default setting at the factory is "dual control" (the distillation endpoint is controlled by time and weight, both using the priority control principle).
4. Distillation experiment begins
First press the 'Confirm' button, then press the heating furnace control button at the bottom left of the fully automatic integrated distillation apparatus to start distillation.