Problem 1: I feel that the repeatability of the sensor force value is not high during use.
solutionUsually, the user needs to handle it personally to confirm whether the signal transmission of the sensor is normal, rule out the possibility of sensor damage, hang a standard weight, such as 10KG, on the sensor end of the tensile machine, and check if the displayed force value has a deviation. If there is a deviation, refer to the manual to enter the calibration program and calibrate it yourself. Some manufacturers' control boards also require a password to enter the calibration program, which requires consulting the manufacturer to provide the calibration password before entering.
Problem 2: The sensor is damaged, replace the sensor.
solutionThe instrument and equipment are broken, which definitely requires the supplier's assistance in handling.
Question 3: Third party correction beyond the legal usage period.
resolvent:This requires a third-party calibration agency to detect force deviation, and the manufacturer or user needs to assist in handling it.