Welcome Customer !

Membership

Help

Fisker Instruments (Shanghai) Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Article

Fisker Instruments (Shanghai) Co., Ltd

  • E-mail

    chinainfo@fesks.com.cn

  • Phone

    13795339399

  • Address

    2nd Floor, Building A, No. 135 Yinyu Road, Jiading District, Shanghai

Contact Now
Comparison and selection suggestions for online detection technology of water alkalinity using titration colorimetric method vs. electrochemical method
Date: 2025-07-25Read: 0
1、 Comparison of operating procedures
Titration colorimetric method
Reagent preparation: It is necessary to prepare hydrochloric acid standard solution, phenolphthalein/methyl orange indicator, and regularly calibrate the reagent concentration (such as 0.1mol/L hydrochloric acid, which needs to be calibrated every 3 months).
Water sample treatment: It is necessary to filter and remove suspended solids (particle size>0.45 μ m) to avoid turbidity and interference with colorimetry; High salinity water samples need to be pre diluted (such as diluting 10 times for seawater alkalinity detection).
Titration operation:
Phenolphthalein alkalinity: Take 100mL of water sample, add 4 drops of phenolphthalein, titrate with hydrochloric acid until colorless, and record the volume V1.
Total alkalinity: Add 3 drops of methyl orange and continue titration until it turns orange red. Record the volume V2.
Result calculation: Total alkalinity (mg/L)=(V1+V2) × C × 50.04/V water sample (C is the concentration of hydrochloric acid).
Operation time: Approximately 15 minutes per test (including reagent addition, titration, and recording).
Electrochemical method (taking electrode method as an example)
Electrode calibration: Use standard buffer solutions (pH 4.01, 7.00, 10.01) for three-point calibration, with each calibration taking 5 minutes.
Water sample collection: No pre-treatment is required (but it is necessary to avoid bubbles adhering to the electrode), and the sample is directly taken to the detection tank.
Real time monitoring: The electrode continuously contacts the water sample and automatically calculates the alkalinity through potential changes (such as pH electrode combined with ion selective electrode).
Data output: The measurement value is updated every 5 seconds and supports 4-20mA signal output to the PLC.
Operation time: 20 minutes for the first calibration, and real-time output for subsequent monitoring.
2、 Selection suggestions
Prioritize the scenario of titration colorimetric method
High precision requirements: For monitoring the total alkalinity of drinking water plants (national standard requirement ≤ 1mmol/L), it must comply with the GB 5749-2022 standard.
Low frequency testing: ≤ 4 tests per day, and the operator has chemical analysis qualifications (such as laboratory environment).
Limited budget: The equipment procurement cost is low (about RMB 20000), but it requires a higher consumption of reagents.
Case: The circulating cooling water system of a certain steel plant adopts MZD-JDS-2012 titrator, which achieves one detection per hour through an automated sampling device, with an error controlled within ± 1.5%.
Prioritize the scenario of electrochemical method
Real time monitoring requirements: such as alkalinity control of aeration tanks in sewage treatment plants (requiring adjustment of dosage every 5 minutes).
Adverse environment: high salinity (such as seawater desalination pretreatment), high turbidity (such as papermaking wastewater) scenarios, electrode method has stronger anti-interference ability.
Long term operation: 24-hour continuous monitoring is required, and maintenance personnel have basic electrical skills (such as electrode replacement).
Case: A wastewater treatment plant in a chemical industrial park uses an EZ alkalinity hardness analyzer to achieve remote alarm through an IoT platform, reducing manual inspection frequency by 70%.
3、 Trends and upgrade directions
Titration colorimetric method: Upgrade to "microinjection+automation", such as using a micro injection pump instead of manual titration, reducing single reagent consumption from 2mL to 0.1mL.
Electrochemical method: Combining AI algorithms to compensate for electrode drift, such as predicting electrode life through machine learning models and triggering maintenance reminders in advance.
Fusion solution: Deploy electrochemical real-time monitoring at key nodes (such as water outlets), and configure titration colorimetric method for regular verification in the laboratory to balance cost and accuracy.