PCT High Voltage Aging ChamberIt is used to test the moisture resistance of products. The tested product is placed under harsh temperature, humidity, and pressure conditions. Moisture will seep into the package along the interface between the gel or gel and the wire frame, multi-layer circuit boards, photovoltaic modules, etc EVA、IC、LED、LCD、 Testing the sealing performance of magnetic materials, polymer materials and other products, testing their pressure resistance and airtightness.
PCT high-voltage aging chamber is suitable for defense, aerospace, automotive components, electronic parts, plastics, magnet industries, pharmaceutical circuit boards, multi-layer circuit boards, etc IC、LCD、 The testing of sealing performance for products such as magnets, lighting fixtures, and lighting products, as well as accelerated life testing for related products, is used in the design phase to quickly expose product defects and weak links. Test the durability and airtightness of its products.
The disposal method of PCT high-pressure aging chamber in case of water shortage:
When the equipment is short of water, the warning system will notify with continuous warning sounds. There are two situations of equipment water shortage: a. The machine has just started and the working time has not started counting, and the water level is insufficient. In this case, the machine cannot start; b、 During the working process, if the water level is insufficient, the machine will sound a warning and cut off the electric heating tube. The temperature will gradually decrease, but the working time will continue.
2. Inject water into the water tank through the water inlet.
3. The work program will automatically continue to execute.
Instructions for using PCT high-voltage aging box:
Before using the PCT aging test chamber, please read the operation and maintenance manual carefully.
2. It is recommended to use distilled water or R.O. water to protect the safety of the pot during use.
3. Please confirm that the manual exhaust valve is closed before using PCT.
4. If the pressure indicator does not return to 0kg/cm after the work is completed?. Before opening the door, please open the manual exhaust valve to eliminate the remaining pressure in the inner pot until the inner pot pressure returns to 0kg/cm?.
After the work is completed, leaving the test items in the pot for too long may cause the pot door to fail to open. This is due to the pressure difference caused by the contraction of hot and cold air. The solution is simple: simply open the manual exhaust valve to remove the gas inside the pot, and the pot door can be opened.