UV aging tester UV aging test chamberIt adopts the original American brandUVThe lamp tube simulates harsh environments such as outdoor sunlight, rainwater, and dew to conduct rapid aging tests on industrial products, so it can also be called ultraviolet radiationUVTester or UV rapid aging chamber. Suitable for aging tests of non-metallic materials against sunlight and artificial light sources; including2Class:UV-AThe wavelength is315nm~380nm,UV-BThe wavelength is280nm~315nm.
UV aging tester UV aging test chamber|ultravioletUVtester|Control system for UV rapid aging chamber:
Taiwan, China programmable7Inch true color touch screen controller(QTEST).
Fixed value operation function.
Users can use this feature to achieve continuous UV testing. In the customized operation screen, users can manually control the opening and closing of various functions. On this screen, temperature setting can be achieved with automatic temperature calculation function.
On the program control screen, users can automatically run the device according to the edited program. When the program is running, the device has functions such as maintaining skip segments. And it can intuitively observe basic information such as the total running time of the program, the running time of program segments, and the program number. At the same time, the operating status of each system will also be displayed intuitively on the screen.
Operation mode: program mode, fixed value mode.
Setting method: Chinese menu, touch screen input.
Interface language type: Chinese/english.
Program editor: editable120Group programs, each group of programs can be programmed100Segment and can be set to run in a loop.
Blackboard temperature monitoring: using standardPt-100Blackboard temperature sensor,Control the surface temperature of the sample during the testing process.
Communication interface:(RS232/RS485Long distance communication1.2km[Fiber optic accessibility30km]).
Auxiliary functions: fault alarm and cause, processing prompt function, power-off protection function, timing function(Automatic start and automatic stop operation)Self diagnostic function.
The black aluminum plate is connected to the temperature sensor, and the blackboard temperature instrument is used to control the heating, making the temperature more stable.
The radiometer probe is fixed and does not need to be installed or removed every time.
Lighting and condensation can be independently controlled and alternately cycled.
The independent control time and alternating cycle control time for illumination and condensation can be1000Set arbitrarily within the hour.
UV tester|ultravioletUVtester|Technical parameters of UV rapid aging chamber:
1.Temperature and humidity range room temperature+10~70℃; Blackboard measurement range:45~80℃; Humidity range: ≥90%RH.
2.temperature fluctuation Volatility: ±3℃; Resolution:0.1℃。
3.Center distance of the lamp 70mm(Center distance between the test sample and the lamp tube: Plate base frame test plate (adjustable up and down).
4.Lamp brand and model the United StatesATLASgenuineUVA-340(315~400nm)TheUVB-313(280~315nm)(Optional), default standard configuration8branch340Lamp tube, if needed313Please explain the lamp tube before placing the order.
5.irradiance 1.0W/m2Internally adjustable(Can output according to the set light intensity).
6.temperature control PIDSelf setting temperature control method.
7.meet the standard GB/T14552-2008The Standards of the People's Republic of China--Plastic, Coating, and Rubber Materials for Mechanical Industry Products - Accelerated Test Method for Artificial ClimateaFluorescent ultraviolet/condensation test method.
lGB/T16422.3-1997 GB/T16585-96Related analysis methods.
lGB/T16585-1996Standard of the People's Republic of China - Test Method for Artificial Climate Aging of Vulcanized Rubber (Fluorescent UV Lamp).
lGB/T16422.3-1997Design and manufacturing standards for corresponding standard clauses such as "Plastic Laboratory Light Source Exposure Test Method".
Meets testing standards:ASTM D4329TheISO 4892-3TheISO 11507TheSAE J2020Waiting for the current UV aging test standards.
Other types of standards are not listed in detail. If necessary, please confirm with our technical department.
About UV tester|ultravioletUVtester|The inner box size of the UV rapid aging box can be selected from standard slicing and non-standard grid layers by Liuqin Technology. For specific solutions and parameters, please contact the manufacturer's "Dongguan Liuqin Testing Instrument Co., Ltd." business department directly or leave a message for consultation.
Next Article: Rain test box Waterproof test box