Date: 2020-08-03Read: 4
  • setThe key has two functions, one is to confirm the configuration, and the other is to press after the configuration is completedsetKey Save Exit
  • exitThe key has the functions of exiting configuration items and canceling configuration
  • timerKey to switch between various options
  • Example: ConfigurationRTCThe Month Power-on startupsetkeyexitkeytimerSwitch to key20setEnter with the keyRTCsubmenutimerThe key is inRTCSwitch from submenu to22setKey to configure (interface flashing) Anup/downSet up
  • Long press during startupsetPress the key to enter the program settings interface

Serial number scope default

01Boiling point setting 90℃--100℃ 100℃

02Low sterilization temperature 105℃--115℃ 105℃

03: Maximum sterilization temperature 140℃--130℃ 135℃

04Low value of dissolution temperature 50℃--70℃ 60℃

05Low value of insulation temperature 40℃--55℃ 40℃

06: Maximum insulation temperature 60℃--80℃ 60℃

07Exhaust temperature value 20℃--40℃ 25℃

08Unlock temperature 50℃--80℃ 80℃

09Sterilization time 1999minute 250minute

10: Insulation time 172hour 4hour

11Dry burning judgment conditions(30second 0to10℃ 2.3℃

20:RTCSettings

21:RTC –year99--00

22:RTC –moon1--12

23:RTC –Sun1-31

24:RTC –time0--23

25:RTC –part0--59

30Calibration Settings

31Calibration settings for steam sensors 129℃--139℃ 134℃

32Temperature sensor calibration settings (item) 129℃--139℃ 134℃

33Calibration settings for water temperature sensor 129℃--139℃ 134℃