Explore HT-350-EXSplit type explosion-proofhumidity meter

Technical Specifications:
Measurement principle:Ceramic moisture sensing
Display mode:4bithighlightLEDdisplay, Displayresolution0.01%
Measuring range:0~40.00%VOL
Measurement accuracy:≤±20FS
heavy again Sex:≤±1.0%FS
Response time:T90<30S
outputinterface:4-20mA. DC(Non isolated output, load resistance<500Ω)
0-5V. DC(Non isolated output, load resistance>)10kΩ)
1Programmable dry contact type passive alarm output,contact point
large capacityAC220V/2A
contact method:RS485(default) orRS232
workpower supplyAC220V±10% 50/60Hz,power consumption<40VA
Instrument operating temperature:-10℃~+50℃
exploreheadWorking temperature: Conventional type:0~200℃
High temperature type:0~400℃
Special type:0~650℃
Sampling method:instrument:Wall mounted, probe:in situformula
prevent protect Sex: Probe Rod316L、instrumentShell protection levelIP65
Probe length:routine0.5mThe1.0mThe1.2mOther lengths can be customized(≤1.8m)
Sensor lifespan:>24Months (under normal usage conditions)
Installation method: plug-in, standardDN65flangeinstall
Product Features:
split-typeDesigned and easy to install,Reduce the impact of external interference on measurement values;
sensor,
highcan650℃Long term use in high-temperature environments within;
haveHigh temperature resistance, high measurement accuracy, weak acid resistance, strong anti-interference ability、Fast response speed and other characteristics;
High precision temperature automatic compensation to eliminate environmental temperatureChanges in measured valuesThe impact;
Two stage filtration design effectively filters dustAnd acidic liquids;
Probe rod316LMaterial, strong corrosion resistance,In harsh environments, other materials can be customized or corrosion-resistant coatings can be sprayed;
Protection level of instrument housingIP65Suitable for use in harsh outdoor environments;
Easy to operate, long service life, and easy to maintain;
Users can freely set4-20mAor0-5VOutput the corresponding humidity upper and lower limits, allowing users to obtain more accurate analog outputs.
Application scenarios:
Widely used in online continuous monitoring system for flue gasCEMSThe measurement of humidity in flue gas from desulfurization and denitrification can also be applied to the humidity control of wood, building materials, papermaking, chemical industry, pharmaceuticals, and tobacco.
Note:If you encounter a low temperature, high humidity, and high dust environment, our company will provide you with a reasonable solution for smoke humidity.
Ordering Instructions(Please specify when placing an order)
lSample gas temperature
lProbe length
Operating parameters (corrosiveness, wind speed, dust content, etc.)