1Six link electric mixerIntroduction:

Six link electric mixerWidely used in laboratories of universities, environmental protection, scientific research, health, petroleum, metallurgy, chemical, medical and other units, it is a commonly used tool for laboratory techniciansThe motor is a permanent magnet DC motor with power enhancement function, stable speed, continuously adjustable, and low noise. It can be stirred simultaneously or separately by six machines, and different speeds can be obtained according to experimental requirements. The stirring rod is made of stainless steel, which enhances the anti-corrosion ability. The biggest feature of this machine is that it is effective when laboratory personnel need slow instrument speed and the viscosity of the laboratory solution is strong for sampling.
IISix link electric mixerPerformance:
1
Electricity Source; single-phase AC50HzThe220V±10%
2
Motor power:60W×6
3Timing range:0-120Minute or normally open
3
, transfer Speed: Start up-3000turn/divide
4Digital display speed,LEDdisplay
5Can be mixed simultaneously or separately
IIISix link electric mixerUsage:
Firstly, check if the accessories are complete. Tighten the mixing rod onto the motor clamp, place the solution bottle on the lifting frame, and adjust the height according to the requirements. The adjustment method is to rotate the handle clockwise and keep both ends raised simultaneously. Connect the power supply, turn on the respective power switches, adjust the speed knob, and select the desired speed from slow to fast. To ensure safety,“Be sure to use a grounding wire'When using, it should be kept clean and dry.
4Six link electric mixerNotes:
1
When using, if you find that the mixing handle is not concentric and the mixing is unstable, please readjust the clamping head to make the mixing concentric. If using a triangular flask for mixing, align the mixing rod with the center and then turn on the machine for mixing.
2
If the power suddenly fails during use, check if the power cord and plug have come off or if the fuse is blown. If there are no such conditions, please notify our factory for repair.