HC-9601Type automatic water quality sampler
an instruction manual
Jintan Jingda Instrument Manufacturing Co., Ltd
Address: Huayang North Road, Jintan Economic Development Zone
1、 Overview:
HC-9601(portableThe mixed sampling automatic water quality sampler is an environmental monitoring instrument launched by our company for sampling surface water and sewage, monitoring water sources, conducting pollution source investigations, and total quantity control. It adopts the commonly used peristaltic pump water intake method and is controlled by a microcontroller. According to user needs, equal proportion mixed water samples and equal time mixed water samples can be collected to meet the sampling requirements in HJ/T91-2002 "Technical Specifications for Surface Water and Wastewater Monitoring" issued by the National Environmental Protection Administration. HC-9601 uses a standard 12V battery for operation, and can be replaced with a car battery when the battery is insufficient for outdoor work. When selecting a flow sensor, the flow is measured simultaneously and the measured flow data and sampling situation are stored. The stored content is not lost in case of power failure and can be accessed afterwards. It is one of the instruments used by environmental monitoring and supervision stations at all levels for water pollution monitoring and environmental research departments for research work.
2、 Main performance indicators:
* Equipped with flow sensor control sampling: 1~9999M ³, arbitrarily set.
* Flow meter control sampling with output pulse: (output pulse is 5V~15V positive pulse). 1-9999 pulses, set arbitrarily.
The flow rate represented by each pulse is set on the flowmeter.
17. Working environment temperature: -5 ℃ to+50 ℃.
18. Power supply: HC-9601 (with built-in DC12V battery)
19. HC-9601Shell protection level: IP65, suitable for use in sewage environments and all-weather conditions.
20. Dimensions: HC-9601(Æ430mm´670mm) (diameter´High),Weight: 23Kg
21. Flow rate: 3.4L/min (when suction is 3m); 2.2/min (when suction is 6m)
III Sampler installation、Precautions for use:
1)Before using the instrument, the battery must be fully charged, otherwise the instrument cannot function properly.
2)The pump tube should be checked regularly, and if it is damaged, it should be replaced immediately. When replacing, there should be a 2mm gap between the pump tube and the pump casing. After replacing the pump tube, press the pump forward or reverse button and let the pump rotate for a while until the pump runs normally without any jamming, and the pump tube is installed.
5)The sampler should be installed above the water intake point and should be avoided from being lower than the water intake point. The water intake pipe should be as short as possible from the sampler to the water intake point, with the end connected to the sampler as the front end and the end at the water intake point as the back end. It should be gradually reduced from the front end to the back end, and there should be no bulge in the middle to avoid water blockage in the tube during reverse blowing of the sampler. Thereby affecting the sampling accuracy.
four. Operation method:
After the system is turned on, enter the following interface:
On the above interface, press (Up key) is for the pump motor to rotate forward, press again (Up key) Stop, press (Down key) Reverse the pump motor and press again (Down arrow) Stop, press (Right click) Enter the manual sampling mode of the instrument, as follows:
|
press (Left click) (Right click) Move the cursor position and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), enter sampling mode, please press to abandonMPress the key to return.
Standby interface, press (Enter key) Enter the automatic sampling mode directly, as follows:
press (Left click) (Right click) Move the cursor position and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), enter automatic sampling mode, please press to abandonMPress the key to return.
Main menu: PressMEnter the main menu with the key,Press again nowMPress the key to exit the main menu and return to the standby interface
press (Enter key) Enter the interface for setting sampling parameters, as follows:
|
|
pressONPress the key to move the cursor down, select the option you want to set, and pressOFFMove the cursor position up with the key, select the option you want to set, and press (Left click) (Right click) Set the required data for individual, ten, hundred, thousand, etc; press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1 If you need to abandon the current settings, simply pressMPress the key to exit and return to the main menu. To save the current settings, please press (Enter key) Save. The pop-up prompt box is as follows:
|
If not covered01Number program, please pressMPress the key to give up, return to the sampling settings interface, re-enter the sampling parameter settings interface, and in the sampling storage number section, enter01Change to other numbers. Additional note: Can be set10Number (program).
Enter the settings query interface as follows:
|
All the numbers (programs) set by the user are here, press (Up key) Number increase1, press (Down arrow) Number digits decrease in descending order1.
pressMPress the key to return to the main menu and enter the flowmeter settings interface, as follows:
|
press (Down arrow key) Move the cursor to the flowmeter interface and press (Enter key) Enter the flow meter selection interface, as follows:
|
Give up, please pressMPress the key to confirm (Enter key) Enter the flowmeter parameter setting interface, as follows:
|
pressONPress the key to move the cursor down, select the option you want to set, and pressOFFMove the cursor position up with the key, select the option you want to set, and press (Left click) (Right click) Set the required data for individual, ten, hundred, etc; press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1 If you need to abandon the current settings, simply pressMPress the key to exit and return to the main menu. To save the current settings, please press (Enter key) Save.
press (Down arrow key) Move the cursor to the interface for selecting pulse flowmeter, as follows:
|
If you need to abandon the current settings, simply pressMPress the key to exit and return to the main menu. To save the current settings, please press (Enter key) Save. The pop-up prompt box is as follows:
|
press (Left click) (Right click) Move the cursor position to set the desired data for one, ten, one hundred, and one thousand, and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Enter the main menu interface and press (Down arrow key) Move the cursor to the system settings and press (Enter key) Enter the clock settings interface, as follows:
|
continue pressing (Enter key) Enter date and time settings:
|
press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1, press (Left click) (Right click) Move the cursor position to set the next set of data.
Enter the pump parameter setting interface, for example: Set the sampling amount to500mlIf higher than500mlReduce the length of the pipe, below500mlAdd a supervisor. press (Enter key) Enter to set the straw length, as follows:
|
press (Left click) (Right click) Move the cursor position to set the desired data for one, ten, one hundred, and one thousand, and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Enter the calibration system settings interface as follows:
|
press (Left click) (Right click) Move the cursor position to set the desired data for one, ten, one hundred, and one thousand, and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Enter the flushing frequency setting interface as follows:
|
press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Move the cursor to password settings and press (Enter key) Enter password modification, as follows:
|
press (Enter key) As shown in the figure, a √ indicates that the password enabled state has been selected, then press (Enter key) Stop password activation. Original password:123456If you need to change your password, enter the password change interface as follows:
|
press (Left click) (Right click) Move the cursor position and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Enter the backlight time setting interface as follows:
|
Note: Backlight time is measured in seconds
press (Left click) (Right click) Move the cursor position and press (Up key) Set the number to increase 1 , press (Down arrow) Set the number to decrease gradually 1Then confirm, please press (Enter key), please press to give upMPress the key to return.
Packing List:
1. 1 host
2. 1 battery
3. 1 charger
4. 1 pump tube (on the pump)
5. 1 water pipe with filter head (6 meters, can be equipped separately according to needs)
6. 1 polyethylene water storage tank (10 liters)
7. 1 copy of instrument user manual