TMP-1500User Manual for Atmospheric Sampler
1、 Scope of use
Jingda Instrument TMP-1500Atmospheric sampler is a commonly used instrument for collecting atmospheric environment and gas in workshop sites. It is widely used in atmospheric environment monitoring, health and epidemic prevention, labor protection, scientific research and other units, and can also be used in conjunction with relevant instruments. By analyzing the sampled gases, we can understand the degree of environmental pollution caused by harmful gases and provide the actual situation of pollution to relevant regulatory authorities in order to take countermeasures and ensure a healthy living environment for people.
2、 Advantages
Jingda Instrument TMP-1500The atmospheric sampler has a compact structure, stable performance, small size, light weight, and is easy to carry, making it an ideal instrument among small portable atmospheric sampling instruments in China. By switching different sampling loads, toxic gases can be collected, as well as dust and activated carbon sampling tubes, making it suitable for multifunctional atmospheric environment monitoring and achieving multiple uses for one machine.
3、 Introduction to Instrument Structure
The main components of the sampler include a diaphragm pump, regulating valve, buffer, flow meter, electronic timing circuit, undervoltage indicator circuit, and power supply.
1.diaphragm pump Its structure and assembly adopt advanced technology, making the small pump have the advantages of compact structure, high efficiency, high pumping pressure, strong load capacity, stable flow rate, low power consumption, and low noise. In100 mmHgUnder load capacity, it is still possible to obtain1L/minabout. The pump valve plate is made of special rubber, which has acid and alkali resistance, aging resistance, and long service life. After long-term testing, its performance still remains good. The pump power motor adopts electronic speed regulation, which can ensure stable operation of the motor even under significant voltage fluctuations and changes in dry battery voltage drops. When the load changes within a certain range during the sampling process, the electronic circuit can adjust the power to stabilize the flow rate of the diaphragm pump.
2.Time control mechanism This machine adopts a fixed type8The timing structure does not need to be set before each sampling, nor will it be unverifiable due to setting errors.
number of files |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
minute |
manual |
5 |
10 |
15 |
30 |
40 |
50 |
60 |
8The gear selection switch can be selected according to the requirements of the conditions, such as the time shutdown in the table, as long as the small switch of a certain gear is turned to“ON'(open)You can obtain the required downtime. Simultaneously shift to second or multiple gears“ON'Timing errors may occur.8The gear switches are all in the off position, and the instrument is not working. Considering that there are different requirements for the selection time of operators“1'The gear is set to manual mode, and users can turn it off according to their own needs.
3.Work instructions and undervoltage indication To ensure the efficiency of operators and make the instrument more practical.TQC-1500The small sampler is equipped with a working indicator and an undervoltage indicator, which can clearly tell the operator when to replace the dry battery. The red light is on during normal operation, and when the voltage drops to6VAt this time, the red light will automatically turn off, indicating that the battery is under voltage. This is also a protective function for nickel barrier batteries.
4.power supply The instrument is equipped with“charge—voltage stabilization'Dual purpose power supply, the power switch position during normal sampling is“voltage stabilization'Position, output at this time8VDC voltage; When used for charging, the power switch is located at“charge'When the position is set and the sampler switch is turned off, it can automatically charge. When the battery voltage of the sampler reaches the rated value, the charger indicator light will automatically turn off, indicating the end of charging.
It is convenient to use dry batteries for outdoor sampling without AC power supply.6section5Dry batteries are generally used4.5About an hour, nickel batteries can be used8About an hour. If the sampler indicator light goes out, it indicates undervoltage and the dry battery needs to be replaced. If the nickel battery is under voltage, it needs to be charged.
4、 Technical parameters
1. Follow upHigh extraction pressure:≥25000pa
2.Rated load flow rate:≥1.5L/min(In40 mmHgtime)
3.Range of traffic usage:0.1~1.5L/min
4.Traffic accuracy:2.5level
5.Stability:≤±5%
6.Electronic timing error:≤±1%
7.Operating Temperature5~40℃
8.Instrument usage status: continuous and intermittent use
9.Power supply: AC/DC dual-use
10.volume137×88×45(mm3)
11.Weight:<0.5Kg(Contains battery)
5、 Usage and maintenance
1.Pull out the battery cover of the sampler and insert it5Dry or rechargeable batteries6Section, cover the cover plate, then gently pull the timing adjustment cover plate to the left and select the desired timing time(for example“2'Location, at this moment5minute)Push the sampler power switch upwards, the red indicator light will turn on, and the instrument will start working. The timing accuracy can be calibrated with a stopwatch.
Attention: If not used for a long time, the battery should be removed to prevent liquid from leaking out and damaging the internal components of the machine when the battery deteriorates.
2.During the sampling process, a filter is connected to filter out impurities such as dust, aerosols, acid and alkali gases in the air; On the other hand, when a backflow of solution occurs, once the solution passes through the filtering fibers of the filter, the fibers will immediately expand, blocking the intake passage and providing protection to the components, especially the pump. In case of solution backdraft, clean the filter and replace it with filter fiber and foam block. If it is used for a long time, the filter fiber and foam block should also be replaced to prevent dirt from passing through the filter mass and entering the pump, which will increase the air resistance and reduce the flow.
3.If accidentally inhaling acidic or alkaline solution during operation, it should be washed several times with distilled water immediately, and then washed with anhydrous ethanol. Generally, the work can be restored. If the damage is severe, it should be sent back to the factory for repair or repaired at the station.
6、 Configuration
1.Instrument accessories
Instrument packaging box(Every two instruments) 1only
Absorption bottle rack1deputy
tripod1deputy
Dual purpose power supply for charging and voltage regulation1only
Dust removal foam plastic block several
filter1only
Instruction Manual1copy