Light incubator is a commonly used experimental instrument, which has certain applications in industrial, chemical, biological, pharmaceutical, scientific research and other fields. The light incubator needs maintenance during use to ensure its normal operation. Today, the editor will introduce the maintenance methods of the light incubator in detail, hoping to help everyone.
Maintenance methods for light cultivation:
1. The outer shell of the incubator should be reliably grounded.
2. The incubator should be placed in a cool, dry, well ventilated place away from heat sources and direct sunlight. Place it steadily to prevent vibration and noise.
3. To ensure effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a 50mm gap on the side of the box, and at least 300mm of space at the top of the box.
4. When handling, repairing, and maintaining the incubator, collision, shaking, and vibration should be avoided; The maximum inclination is less than 45 degrees.
5. The instrument suddenly doesn't work. Please check if the fuse tube (behind the box) is burnt out and check the power supply.
6. When cooling the incubator, it is not advisable to make the temperature difference between the incubator and the ambient temperature greater than 25 degrees.