Welcome Customer !

Membership

Help

Changzhou Guowang Instrument Manufacturing Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Download Materials>Detailed Technical Information on JWL-2 Secondary Air Microbial Sampler
Download

Changzhou Guowang Instrument Manufacturing Co., Ltd

  • E-mail

    jtgwyq@126.com

  • Phone

    13511666605

  • Address

    D16, No. 39 Yangang East Road, Jincheng Town, Jintan District, Changzhou City, Jiangsu Province

Contact Now
Detailed Technical Information on JWL-2 Secondary Air Microbial Sampler
Date: 2013-08-10Read: 0

Our factory specializes in productionJWL-2 Secondary Air Microbial SamplerOur manufacturer specializes in the quality and largest scale of air microbial samplers The service Feedback to customers, inquiring about air microbial samplers and other products, air microbial samplers have always received positive feedback from the market.

1JWL-2 Secondary Air Microbial SamplerCharacteristics:

The serious harm caused by air microbial pollution has received increasing attention today, making the need for sampling and monitoring of various air pollutants even more urgent. The quantity and size distribution of air microorganisms are two * indicators for evaluating the hazards. The JWL-2 air microorganism sampler produced by our factory can not only measure the quantity of air microorganisms, but also the size of these particles, which is one of the important indicators for determining the harm of air microorganisms. He is composed of six machines combined into one, and each stage is actually a single-stage sampler. Using the principle of two repeated impacts, the vast majority of particles, especially those settling in organs and lungs, are basically impacted. Therefore, the particle size range it collects is naturally wider than that of single-stage impact samplers. The circular nozzle of the impactor has a higher sampling efficiency than the slotted nozzle. The relative humidity gradually increases during sampling, which is very beneficial for the survival of fragile pathogenic microorganisms, especially virus particles, and has been commonly used since its inception.

IIJWL-2 Secondary Air Microbial SamplerPurpose:

JWL-2 Secondary Air Microbial SamplerIt can be widely used for sampling and monitoring the quantity and size distribution of air microorganisms in environments such as disease prevention and control, environmental protection, pharmaceuticals, fermentation industry, food industry, and biological cleanliness. It can also be used for sampling and research on air microorganisms in relevant scientific research and teaching departments, providing scientific basis for evaluating the hazards of environmental air microbial pollution and its control measures.

IIIJWL-2 Secondary Air Microbial SamplerTechnical Specifications:

1. Capture rate:>98%

2. Capture particle range:

*Grade>7.0 um aperture 1.18mm

Second level 0.65 μ m-1.1 μ m aperture: 0.25mm

3. Sampling flow rate 28.3 L/min (adjustable)

4. Power supply: AV220V Power: 35W

5. Volume: Impactor: ⊙ 108mm × 82mm Host: Length 200mm × Width 150mm × Height 125mm

6. Weight: Impactor: 380g Host: 3Kg

4JWL-2 Secondary Air Microbial SamplerCultivate and count bacterial colonies:

1. Invert the sampled petri dish and incubate it in a 37 ℃ constant temperature incubator for 48 hours. For microorganisms with special requirements, incubate them under corresponding conditions.

2. Count the number of colonies on each level of petri dish, where each colony is a colony forming unit (CUF).

FiveJWL-2 Secondary Air Microbial Samplerto configure:
1. Host: One set (including one vacuum pump, one flow agent, and one timer each)
2. Impactor: One Level 2 Impactor
3. Tripod: one unit
4. Operation manual: one copy
5. Accessories such as connecting pipes: one set
6. Aluminum alloy suitcase: one