Welcome Customer !

Membership

Help

Beijing Zongheng Jinding Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Video
Video

Beijing Zongheng Jinding Instrument Equipment Co., Ltd

  • E-mail

    18612919058@163.com

  • Phone

    18612919058

  • Address

    No. 86 Wanxing Road, Changyang, Fangshan District, Beijing

Contact Now
Brief Introduction

LJC-80KV typeInsulation oil dielectric strength tester

1. Introduction
The dielectric strength tester for insulating oil is based on international standardsIEC-156In accordance with national standardsGB507-86According to the requirements of the "Method for Measuring the Dielectric Strength of Insulating Oil" and feedback from usersHcj-9101On the basis of this, a fully automated instrument has been developed.
This instrument uses a microcontroller as the main component and adopts a method of setting up first and then starting up for testing. The entire process is automatically controlled by the microcomputer, which is easy to operate and convenient to use.
2、 Purpose and Characteristics
This instrument is suitable for testing the dielectric strength of various insulating oils. Its main performance characteristics:
1.This instrument is equipped with an automatic detection function, which automatically enters the reset state when turned on and executes the regulator to return to zero.
2.Adopting micro technologyTPU-APanel printer with automatic printing output.
3.According to user needs, the number of tests, mixing and settling time, and sound and light control reminders for continuous printing and non printing can be changed.
4.This instrument adopts fully automatic magnetic oscillator stirring to eliminate unevenness and bubbles in the oil sample.
3、 Technical specifications
1.Working power supply:AC220V ±10%The50HZ±5%
2.Measuring range:AC 0-80KV
3.Limited current:5mA
4.Measurement accuracy:1%
5.Pressure regulation speed:2KV/S±10%
6.Reservation settings: number of times1-9
mixing time 0-1divide39second
standing time 0-10divide39second
7.Usage conditions: ambient temperature0℃-35℃
relative humidity≤75%
8. Oil cup gap:2.5mm(Diameter of oil cup plug gauge)
9. volume: 415×315×315mm3
10. heavyQuantity:28kg