The accuracy of the LDH-15KV arc resistance tester should pay attention to the following points:
Equipment calibration and maintenance
Electrode statusRegularly clean or replace worn electrodes (such as tungsten electrodes) to ensure precise electrode spacing (e.g. 6.35 ± 0.1mm) and avoid electrode erosion affecting discharge stability1
2
.
High voltage system calibrationCalibrate the voltage output every six months, with an error controlled within ± 2%, and equip a voltage regulator to ensure stable input voltage1
.
Gas systemControl the gas flow rate (usually 10L/min) and purity, check the sealing of the pipeline to prevent gas leakage from affecting the test results1
.
Sample preparation and installation
Size and processingProcess samples according to standards (such as a diameter of 50mm and a thickness of 3mm), polish the surface and clean it with anhydrous ethanol, and pre treat for 24 hours (under conditions of 23 ± 2 ℃ and 50 ± 5% humidity)1
2
.
Installation specificationsFix the sample with a fixture to ensure that it is perpendicular and in close contact with the electrode, avoiding testing errors caused by poor contact1
.
Parameter Setting and Operating Standards
parameter matchingAdjust the voltage (10-15kV) and current (such as 10/20/40mA) according to the material characteristics. The new material needs to optimize the parameter combination through pre testing1
2
.
environmental controlMaintain the humidity of the testing environment between 40-60% to avoid contamination of the sample surface by oil or dust1
.
Operational SecurityWear protective equipment during operation, close the test chamber door, and avoid contact with high-voltage parts1
3
.
Data recording and maintenance
real-time monitoringRecord the failure time of the sample (such as carbonization or combustion), use the system to automatically collect data, and ensure the accuracy of the data1
.
regular maintenanceAfter every 50 tests, check the electrode status, clean the carbide residue, and calibrate the equipment1
.
By strictly following the above precautions, the testing accuracy of the arc resistance tester can be effectively improved, ensuring the reliability and consistency of the test results.