Welcome Customer !

Membership

Help

Beijing Xuxin Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Article

Beijing Xuxin Instrument Equipment Co., Ltd

  • E-mail

    zhangq@bj-xuxin.com

  • Phone

    13552591542

  • Address

    Shanshui Taihe Entrepreneurship Park, Shahe Town, Changping District, Beijing

Contact Now
What are the calibration aspects of mechanical impurity analyzer
Date: 2025-12-11Read: 1
The calibration of mechanical impurity analyzer is the core link to ensure the accuracy and reliability of detection results, and it is necessary to strictly follow national standards (such as GB/T 511) and instrument characteristics for systematic operation. The following will explain from three aspects: calibration process, key parameter verification, and maintenance management:
1、 Preparation before calibration
-Environmental condition control
The instrument should be placed in a laboratory without vibration or strong electromagnetic interference, with an ambient temperature of 15 ℃ to 35 ℃ and a relative humidity below 85%. Avoid direct sunlight or proximity to high-temperature equipment (such as ovens) to prevent plastic parts from aging or affecting temperature control accuracy.
-Preparation of standard materials and equipment
-Standard weights: used for verifying the linearity, repeatability, and sensitivity of analytical scales.
-Quantitative filter paper/glass sand core funnel: It needs to be dried at 105 ℃~110 ℃ in advance to a constant weight (difference between two weighings ≤ 0.0004g) to ensure the initial mass stability of the filter medium.
-Standard sample: Select quality control samples with known impurity content to verify the overall performance of the instrument.
-Initial inspection of instrument functions
After powering on, check the touch screen display, button response, and vacuum pump negative pressure capability (usually up to -70 kPa). Confirm that the heating device (such as constant temperature funnel, oven) is working properly, and the temperature fluctuation range should meet the standard requirements (such as ± 1 ℃).
2、 Calibration methods for key parameters
-Temperature control system
-Constant temperature funnel and oven: Insert a precision thermometer into the funnel bath or place it in the center of the oven, set a common temperature point (such as 105 ℃), and compare the displayed value with the measured value after stabilization. If it exceeds the allowable range (such as ± 2 ℃), the temperature control module needs to be adjusted or the sensor needs to be replaced.
-Furnace plate heating: Similar methods are used to verify its uniformity, avoiding local overheating that may cause carbonization of filter paper or decomposition of impurities.
-Vacuum system
Connect a calibrated vacuum gauge and monitor the negative pressure stability during the filtration process (usually ≤ -0.08MPa). Excessive pressure fluctuations can cause changes in filtration efficiency, and it is necessary to check the sealing of the pipeline (such as aging rubber pipes) or the performance of the pump body. If air leakage is found, the sealing gasket should be replaced in a timely manner.
-Correlation between timing and weighing
-Timing function: Compare the time error of the filtration and drying steps through a stopwatch to ensure compliance with standard regulations (such as controlling the filtration time within 5 minutes).
-Weighing accuracy: Use standard weights to calibrate the analytical balance, requiring an error of ≤± 0.0002g. Filter paper constant weight and sample weighing must be completed on the same calibrated balance to avoid system errors caused by equipment differences.
3、 System performance verification and actual operation calibration
-Standard sample testing
Conduct full process determination of samples with known impurity content according to the GB/T 511 process, and compare the relative deviation between the measured value and the nominal value. If the result exceeds the limit (such as repeatability error>0.002%), it is necessary to investigate the filtration efficiency, drying effect, or human operation factors.
-Consistency verification of solvent and filter material
When replacing solvents or filter membranes from different batches, it is necessary to conduct a blank test again to eliminate interference from the background of the medium. For example, when using a benzene ethanol mixture, it should be ensured that its purity meets the standards to avoid residual impurities affecting the results.
4、 Calibration cycle and record management
-Calibration frequency
Perform comprehensive calibration before use, after major repairs, or at least once a year. In daily use, it is necessary to use standard weights to check the status of the balance every month, and to check the quality of the vacuum pump oil (clear without emulsification) every quarter.
-Record archiving
Detailed records of calibration date, environmental conditions, standard information, measured data, and adjustment measures. For example, "On December 10, 2025, calibrate the oven and set the temperature to 105 ℃. The measured temperature was 106.5 ℃, with a deviation of+1.5 ℃. The sensor has been calibrated. This type of record provides a basis for quality traceability.
The calibration of mechanical impurity analyzer needs to run through the entire chain of "temperature vacuum time weighing", combined with standard substance verification and preventive maintenance. A standardized calibration process can not only enhance the reliability of detection data, but also extend the lifespan of instruments and reduce the risk of cross contamination.