KTThe series of low-frequency neuromuscular stimulators consists of two models, KT-1 and KT-2,KT-2It is a new model product derived from our company's KT-1 by adding inter channel pulse output delay.
KT-2 Low Frequency Neuromuscular Stimulator (Spasmodic Muscle Treatment Device)It uses the principle of low-frequency pulses to act on human soft tissues, stimulate muscle contraction, and achieve the goal of reducing and relieving pain caused by soft tissue injury.
KT-2 type neuromuscular stimulator (spastic muscle therapy device)Designed for the physiological and pathological characteristics of neuromuscular tissue using low-frequency pulse electrical therapy. Selective application of asymmetric bidirectional pulse waves with different frequencies and pulse widths to the spastic and antagonistic muscles of patients, alternating between the two to stimulate muscle contraction, achieving the goal of relaxing muscles, relieving muscle spasms, and improving limb function.
This stimulator has four treatment channels and can be used for two people simultaneously.Channel I and Channel II treat one patient together, while Channel III and Channel IV treat the otherCan treat another patient simultaneously. Multi site treatment can also be performed on a single patient. The method is skin electrodes (which can also be used for electroacupuncture stimulation).
This product consists of a host, wires, and electrode pads.
This product has analgesic and anti-inflammatory effects on bone hyperplasia, lumbar muscle strain, torsion and contusion, joint pain, and has a stimulating effect on muscles, reducing muscle tone. But the body is prone to adaptability.
Acute suppurative inflammation, bleeding tendency, malignant tumors, thrombophlebitis, active pulmonary tuberculosis, patients with pacemakers, pregnant women, local foreign bodies, heart area, lower abdomen of pregnant women, and those who cannot tolerate electrical current.
1Environmental temperature: 5 ℃ to 40 ℃.
2Relative humidity: ≤ 80%.
3Atmospheric pressure: 500hPa to 1060hPa.
4Power supply conditions: AC 220V ± 22V, 50Hz ± 1Hz.
5Input power: 20VA.
1Pulse frequency:0.5Hz~1HzContinuously adjustable (pulse interval)T:1S~2S,step0.1SContinuously adjustable)(500Ω), allowable deviation error ±15%.
2、Pulse width: 0.1~1ms,step0.1msContinuously adjustable(500Ω),tolerance±30%.
3、Each pulse power: When the output amplitude is large, the power of each pulse should be greater than7цC.(500Ω)
4Single pulse energy:When the output pulse width is less than 0.1s, the energy output of a single pulse does not exceed 300mJ.(500Ω)
5Peak open circuit output voltage: not greater than 500V.
6Delay time: Channel II output pulseT2Output pulse of channel IT1(Or Channel IV output pulseT4Compared to the output pulse of channel IIIT3)Delay occurs. delay timeTdelay0.1~1.5s,step0.1sContinuously adjustable, with a tolerance of ±15%.(500Ω)
7Treatment time: divided into5min~60min,part12Block, with an error of±10%.
8The stimulator should be able to withstand the effects of open and short circuits at the output end without compromising its performance.
9When the power supply is interrupted and restored, there should be no output.
10、Output current: The effective value of the large output current should not exceed 80mA.(500Ω)
11Output amplitude adjustment: Each increment should not exceed 1mA or 1V for discrete increases, and small output increments should not exceed 2% of large outputs.
12Output mode: continuous operation.
13Output waveform: Pulse wave.
14Continuous working time: ≥ 4h.
15Fuse: F2A250V, Φ 5 × 20mm.
16Safety type: Class I BF type.
17Work system: continuous operation.
18Dimensions: 490mm × 400mm × 200mm.
19Weight: 11kg.
20Number of output channels: 4; Each channel is independently set and connected to two electrode pads.