- Phone
-
Address
601, 6th Floor, Building 1, No. 7 Xijing Road, Shijingshan District, Beijing
Beijing Ruizhongcheng Trading Co., Ltd
601, 6th Floor, Building 1, No. 7 Xijing Road, Shijingshan District, Beijing
It uses the principle of low-frequency pulses to act on human soft tissues, stimulate muscle contraction, and achieve the goal of reducing and relieving pain caused by soft tissue injury.
DXZ-2 Computer Low Frequency Diagnosis and Treatment DeviceIt applies low-frequency pulse currents to peripheral nerves and muscle tissues, and the instrument determines treatment methods and rehabilitation levels based on the diagnosis of peripheral nerve and muscle tissue diseases.
This product combines diagnosis and treatment, is easy to use, and uses skin electrodes for treatment.
This product has a strong stimulating effect on sensory and motor nerves. It has therapeutic and adjuvant therapeutic effects on diseases such as bone hyperplasia, lumbar muscle strain, lumbar muscle fibrosis, tennis elbow, torsion and contusion, joint pain, etc.
This product consists of a host, wires, and electrode pads.
This product can diagnose and treat muscle condition, and has analgesic and anti-inflammatory effects on bone hyperplasia, lumbar muscle strain, torsion and contusion, joint pain, lumbar muscle fibrous tissue inflammation, and tennis elbow. But the body is prone to adaptability.
Acute suppurative inflammation, bleeding tendency, malignant tumors, thrombophlebitis, active pulmonary tuberculosis, patients with pacemakers, pregnant women, local foreign bodies, heart area, lower abdomen of pregnant women, and those who cannot tolerate electrical current.
1Environmental temperature: 5 ℃ to 40 ℃.
2Relative humidity: ≤ 80%.
3Atmospheric pressure: 500hPa to 1060hPa.
4Power supply conditions: AC 220V ± 22V, 50Hz ± 1Hz.
5Input power: 40VA.
1Pulse frequency: 0.2-1000Hz (under 500 Ω conditions), divided into 197 levels, with an error of ± 15%.
2Pulse width: 0.01ms~1000ms (under 500 Ω conditions), divided into 207 levels, with an error of ± 30%.
3Each pulse power: When the output amplitude is large, the power of each pulse should be greater than 7 μ C (under 500 Ω conditions).
4Output current: The effective value of the large output current should not exceed 80mA (under 500 Ω conditions).5Single pulse energy: When the pulse width is less than 0.1s, the energy output of a single pulse does not exceed 300mJ (under 500 Ω conditions).
6Peak open circuit output voltage: not greater than 500V.
7The diagnostic and therapeutic device should be able to withstand the effects of open and short circuits at the output end without compromising its performance.
8When the power supply is interrupted and then restored, there should be no output.
9Output amplitude adjustment: Each increment should not exceed 1mA or 1V, and the discrete increase should not exceed 2% of the large output increment.
10、Amplitude modulation rhythm: 2 to 40 beats per minute, divided into 39 levels, with an error margin of ± ±10%(500ΩUnder certain conditions).
11、Treatment time: 5 minutes~60min,Divided into 12 levels,Error±10%.
12Single pulse output should have a single pulse output mode (500ΩUnder certain conditions.
13Output waveforms: square wave, triangular wave, sawtooth wave, trapezoidal wave, step wave, exponential wave, sine wave, DC.
14、output methodEqual amplitude and amplitude modulation.
15Number of output channels: 1; Connect two electrode plates.
16Continuous working time: ≥ 4h.
17Fuse: OutputF2A250VΦ 5mm × 20mm. InputF0.1A250VΦ5mm×20mm
18Security type: Type I, BF application part.
19Operating mode: Continuous operation.
20Dimensions: 500mm × 420mm × 970mm.
21Weight: 11Kg.