The latest upgraded version of hydrogen fluoride online conductivity pool is online
Date: 2022-10-13Read: 28

Based on feedback from different users and practical situations,Further upgrades and improvements have been made to the online conductivity cell for hydrogen fluoride, while meeting the needs of both production sites and laboratories.

The integrated PTFE lining inside the pipeline can withstand long-term corrosion from hydrogen fluoride or hydrofluoric acid solution.

Built in high-purity Pt gold electrode, external copper electrode interface (4mm socket); Fixed with flanges for easy disassembly and maintenance.

Built in isolated temperature probe to prevent corrosion from liquid materials; The probe specifications can be customized according to the requirements of the conductivity meter.

Can adapt to various specifications of conductivity meters.

Special requirements can be negotiated and customized.

Inner diameter of pipeline: 28mm; length of tank body: 280mm

Electrode specifications: high-purity platinum, electrode diameter 2mm, spacing 100mm

Temperature detection: Pt100 isolated probe (not corroded by the material liquid) (other specifications of temperature probes can also be equipped according to the requirements of the conductivity meter)

Flange specifications: DN25 or DN50

Installation method: Online