E-mail
lihuizhi@accu-flow.com.cn
-
Phone
15010116300
-
Address
Building B, Ruichuang International Center, No. 8 Wangjing East Road, Chaoyang District, Beijing
Beijing Jingliang Technology Co., Ltd
lihuizhi@accu-flow.com.cn
15010116300
Building B, Ruichuang International Center, No. 8 Wangjing East Road, Chaoyang District, Beijing
11.29,20220
Equipped with a display screenControl video of Coriolis/Coriolis mass flow controller/flowmeter:
The table body itself has a display screen and control buttons;
◆ Can directly measure mass flow rate
The accuracy can reach ± 0.5% F.S for gases and ± 0.25% for liquids
◆ Fast response speed and fast adjustment speed
High precision and good repeatability
◆ Can simultaneously output fluid density and temperature
High viscosity fluids and high-density gases can be measured
◆ Not significantly affected by environmental vibrations
The medium of the flow controller in this video is carbon dioxide, and the pressure is 20Mpa. You can see the control effect after setting different flow values.
(Some parameters are as follows. If there are special requirements, they can be customized)
| measurement range | 20g/h~500kg/h |
| Measurement and control range | Flow meter range ratio 100:1 Controller range ratio 50:1 |
| Accuracy (Full Scale) | Gas: ± 0.5% F.S, liquid ± 0.25% F.S |
| stability | ±0.1%F.S |
| repeatability accuracy | ±0.05%F.S |
| response time | <0.2s |
| digital quantity | RS232/485, MODBUS protocol |
| analog quantity | 0~5V、4-20mA、1~5V |
| power supply | ±15VDC,24VDC |