Portable crude oil water content analyzer model: HAD-29624
1. Overview of HAD-29624
The HAD-29624 portable crude oil moisture analyzer (hereinafter referred to as the instrument) is a compact instrument used to measure the moisture content in crude oil, finished oil, and other chemical solutions. The instrument is designed to be lightweight, precise, stable and reliable, easy to use, and has fast measurement speed. Suitable for sampling wellhead pipelines and laboratories in the petrochemical industry, measuring the moisture content of finished oil, and measuring the moisture content of various recovered oils.
2. Characteristics of HAD-29624 surgery
2.1 Lightweight and easy to carry: The total weight is only 1kg.
2.2. Strong capabilities: Sensor data acquisition, software processing and display, computer debugging software for easy calibration by users.
2.3. Fast measurement speed: It only takes a dozen seconds from sampling to measurement.
2.4 Measurement accuracy: The digital sensor used has resolution, stability, reliability, and accuracy.
LCD can display up to two decimal places (00.00%).
2.5. Low consumption: Powered by a rechargeable lithium battery of 0.15W. It can detect nearly 300 oil samples when fully charged.
2.6 Convenient operation: Press the power switch during measurement, wait for a few seconds for the percentage to be displayed, insert it into the oil sample to be tested, and read the data displayed on the LCD.
3. Composition and working principle of HAD-29624
Composition: This instrument consists of a resolution digital sensor and a microcontroller for data acquisition and processing.
Working principle: The portable water content detector detects water content by generating different resonant frequencies in different media through high-frequency short waves. When the sensor is inserted into pure oil, the output voltage value is small; The voltage value output when the sensor is inserted into pure water; When the sensor is inserted into an oil-water mixture, the output voltage will display the corresponding water content from small to large on the screen due to the proportion of oil-water mixture.
4. Main surgical indicators
4.1 Measurement range: 0-60% ()
4.2. Degree: 0-60%: ± 0.1% -0.5%; (60%-:±5%)
4.3 Resolution: 0.01%
4.4 Power supply: DC9V lithium battery
4.5. Temperature range: -20 ℃ -85 ℃
4.6. Humidity: ≤ 85%
4.7. Dimensions: 330mm × 70mm × 26mm (length, width, thickness)
4.8 Quality: not exceeding 1kg