Instrument configuration
GS-MS:6890N+5973
Pre processing equipment:Steady DeYan AutoTDS VPro Fully Automatic Thermal Desorption Instrument
Adsorption tube activatorPractical DeYan BTH-24 Activation Instrument
Adsorption tube:TenaxTAAdsorption tube
AutoTDS-VProType fully automatic thermal analyzer

AutoTDS-VProThe Type fully automatic thermal desorption instrument is a 24 bit low-temperature secondary fully automatic thermal desorption instrument with electronic flow control (EPC), which has a diversion function and can store multiple operating methods. It can automatically complete 24 adsorption tubes.
BTH-24Type Activation Instrument

BTH-24The Type Activation Instrument has a simple appearance, built-in flow sensor, independent temperature control system, and 24 adsorption tubes divided into two groups for independent control. It is easy to operate, flexible and versatile, and can meet various needs.
Instrument conditions
GCMScondition
Injection port: Temperature: 200 ℃, Pressure: 4.77Psi, Split ratio: 5:1, Split flow rate: 4ml/min
Chromatography column: AB-5MS, 30m * 0.25mm * 0.25um, column flow rate 0.8ml/min
Program heating up:

Thermal desorption conditions

Test results
22Total ion chromatogram of TVOC (100ng):

1-N-hexane, ethyl acetate, 3-3 chloromethane, 4-phenyl, 5-carbon tetrachloride, 6-cyclohexane, 7-n-heptane, 8-trichloroethylene, 9-cyclohexane, 10 toluene, 11 octane, 12 tetrachloroethylene, 13 butyl acetate, 14 chlorobenzene, 15 ethylbenzene, 16 para, meta xylene, 17 styrene, 18 o-xylene, 19-n-hexane, 20-1,4-dichlorobenzene, 21-n-hexadecane
22The peak of the target component is normal. Tenax adsorption tubes with component contents of 25ng, 50ng, 100ng, 200ng, 500ng, and 1000ng were prepared using a preparer. The correlation coefficients R ² of each component were all greater than 0.999, meeting the standard requirements