E-mail
2212240822@qq.com
-
Phone
18911395947
-
Address
514, Unit 1, Building 2, Xingfu Square, Gu'an County, Langfang City, Hebei Province
Beijing Beiguang Jingyi Instrument Equipment Co., Ltd
2212240822@qq.com
18911395947
514, Unit 1, Building 2, Xingfu Square, Gu'an County, Langfang City, Hebei Province
This instrument adopts the principle of ultraviolet oxidation to oxidize organic matter in the sample into carbon dioxide. The direct conductivity method is used to test the carbon dioxide content. The total organic carbon content is determined by measuring the difference between the total carbon content of the sample that has undergone oxidation reaction and the total inorganic carbon content of the sample that has not undergone oxidation reaction, that is, total organic carbon (TOC)=total carbon (TC) - total inorganic carbon (TIC).
Product Features
1. The instrument adopts a portable design, which is easy to use and convenient to move to the sampling point.
2. Adopting an embedded system and touch screen design, pure Chinese operation is convenient and easy.
3. Design and conduct testing for the total organic carbon content of pharmaceutical water (TOC content below 1000ppb).
4. Equipped with a large amount of storage space, capable of storing a large amount of test data.
5. Print in Chinese and output test parameters and results.
6. No gas or reagents are required during use, storage, and replacement, and there are no moving parts, reducing maintenance and repair costs.
When the concentration of the test sample exceeds the specified limit, the instrument can automatically sound an alarm and output a control signal.
8. Testing plans that comply with the requirements of the Chinese Pharmacopoeia can provide IQ/OQ/PQ services.
Performance specifications:
Measurement range: 0.001mg/L~1.0mg/L (sensor can be customized, concentration can be adjusted up to 1000mg/L, sensor can be customized and adjusted to a certain concentration range according to the style requirements)
Accuracy: ± 4% Test range
Resolution: 0.001mg/L
Analysis time: continuous analysis
Response time: within 4 minutes
Detection limit: 0.001mg/L
Sample temperature: 1-95 ℃
Repetitive error: ≤ 3%
Power requirement/function: 220V
Display screen: Color touch screen