Volume surface resistivity tester
Product model: BEST-212
Instruction Manual
*(Please read the user manual carefully before use)*
Beijing Beiguang Jingyi Instrument Equipment Co., Ltd
Declaration
Thank you for choosing Beiguang Precision Instrument! It is our great honor to have you as our customer. Our company not only provides you with high-quality products, but also offers you satisfactory services! In order for you to use this experimental instrument more proficiently, we have randomly provided this manual.
This manual mainly introduces the knowledge related to the structural principles, operating methods, and safety precautions of this equipment.
Special reminder: Sometimes, in order to improve the performance of the device, we may make some modifications to the software or mechanical parts, which may result in inconsistencies between the operating system and the user manual in certain details. Hereby declare: The random instruction manual of the equipment you purchased is subject to the actual equipment of the testing instrument. When writing this manual, we inevitably have errors and omissions. Please forgive us and warmly welcome your valuable opinions or suggestions.
The content of this manual is subject to change without prior notice.
This manual is copyrighted by Beijing Beiguang Jingyi Instrument Co., Ltd; Without written permission from Beiguang Jingyi Instrument Co., Ltd., no part of this manual may be photocopied, copied, or translated into any other language in any way.
Please pay special attention to the areas marked with "★" in this book and read them carefully.
Random file:
Randomly install the following files:
1. One copy of the user manual
2. One copy of the 'Product Factory Packing List'
3. One copy of Product Warranty Card and Certificate of Conformity
When you receive this device, please check if the above documents are complete and keep them properly.
★ Special statement: Depending on the specific requirements of the customer, please refer to the packing list for specific configurations.
This manual cannot be used as a basis for making any requests to our company.
The interpretation of this manual belongs to our company.

catalogue
matters needing attention. ............................................................................................... . .................................... 1
Operation method ............................................................................................... . ....................................2
Overview: ...................................................................................................................... 3
Two main characteristics ................................................................................................................................. 3
Three main application areas .......................................................................................................................... 3
Four technical indicators ...................................................................................................................................3
Five working principles ...................................................................................................................................4
Six usage methods ................................................................................................................................... 5
6.1. Connect the power cord properly ............................................................................................................................ 5
6.2 Connect the power supply ...................................................................................................................................5
6.3 Zero adjustment ...................................................................................................................................... 5
6.4 Connecting Lines ................................................................ . ..................................................................5
6.5 Choose the appropriate measurement voltage ..................................................................................................................5
6.6 Testing ............................................................................................................................................5
Seven precautions for use .............................................................................................................................5
7.1 Zero adjustment should be made when there is an open circuit at both ends of "Rx" ..................................... . ...............................................................5
7.2 It is prohibited to short-circuit the "Rx" terminals to prevent the microcurrent amplifier from being impacted by high currents ................................................5
During the testing process, do not arbitrarily modify the measured voltage .......................................................................................5
During measurement, gradually shift from low gear to high gear ........................................ . ....................................................5
The resistance value of most insulation materials will change to some extent after voltage is applied, causing numerical changes .....................................6
After connecting the power supply, fingers should not touch the metal parts of the high-voltage line .......................................................................6
During the 7.7 testing process, it is not allowed to interact with the microcurrent testing terminal ..............................................................................................6
When measuring high resistance, a shielding box can be used to shield the object being measured .......................................................................6
7.9 After each measurement, the range switch should be turned back to the "104" position before the next test ..................................................6
Eight typical applications ....................................................................................................................................6
8.1 Measure the resistance value of anti-static shoes and conductive shoes ................................................................................................6
8.2. Measure the resistance and electrical resistivity of anti-static materials ...............................................................................................6
8.3. Measure the system resistance value of the raised floor used in the computer room ............................. . ...................................................7
8.4. Measure the resistance (rate) of insulation materials ...............................................................................................................7
8.5. Measurement of current and ultra-high resistance above 1015 Ω ..................................................................................7
9、 Common problems ....................................................................................................................................7
Why do different resistance ranges have different readings when measuring the same object .....................................................7
Why do different measurement ranges still produce significant differences in the results when measuring the current of some objects? ....................... .7
Why is the reading of the instrument always unstable during measurement ................................................... . ......................................7
Why is it necessary to turn the range switch back to 104 before turning off the power after the measurement is completed? ......................................7
Why not change the test voltage of the tested object during the resistance measurement process? ............................. . .......................7
Why do I have to turn off the power after setting the range switch to 104 after completing the measurement? ..........................................7
10、 Three electrode wiring instructions .....................................................................................................................8
11、 Instrument maintenance ...............................................................................................................................10

matters needing attention
★★★★★★★★★★★★★
Please read the following carefully before using the instrument, otherwise it may cause damage or electric shock to the instrument.
1. ★ Check if the voltage range on the back panel of the instrument is set to 10V and if the current resistance range is set to 104.
2. ★ Turn on the power and zero it (note that the host should not be connected to the shielding box circuit at this time). In the case of an open circuit at both ends of "Rx", zero it to make the ammeter display 0000. Then shut down.
3. ★ Lay the test sample flat in the center of the unprotected electrode, then press the sample with the protective electrode, and insert the protected electrode (unprotected electrode, protective electrode, and protected electrode should be coaxial and confirm that there is no short circuit between the electrodes).
When measuring volume resistance, turn the test button to the Rv side, and when measuring surface resistance, turn the test button to the Rs side,
5. ★ Connect the test cable and connect the host to the shielding box. Set the range to 104 and turn on the power switch button on the back panel of the host. Adjust the voltage button on the back panel of the instrument to the required measurement voltage. (For example, in the GBT 1692-2008 standard for measuring the insulation resistivity of vulcanized rubber, it is specified that the measurement should be carried out at a voltage of 500V, so the voltage should be raised to 500V.)
6. ★ Gradually shift the current resistance range button from low to high, hold for 1-2 seconds each time, and observe the displayed number. When the measured resistance is greater than the instrument's measurement range, the resistance meter will display "1". At this time, continue to move the instrument to a higher range position. When the measuring instrument displays a value, it should stop and measure the resistance after 1 minute of electrification time. The current number multiplied by the level is the measured resistance.
7. After the test is completed, first set the range to (104), then set the measured voltage to 10V, and finally turn off the power after setting the test button to the center position. Then proceed to the next test.
8. ★ It should be zeroed when there is an open circuit at both ends of "Rx". Generally, it is not necessary to zero again during the testing process after one zeroing.
9. ★ It is prohibited to short-circuit the "RX" terminals to prevent the microcurrent amplifier from being impacted by high currents.
10. ★ Do not arbitrarily change the measured voltage during the testing process.
11. ★ Gradually shift from low gear to high gear during measurement.
12. ★ After connecting the power, fingers should not touch the metal parts of the high-voltage line.
13. ★ Do not touch the microcurrent testing terminal during the testing process.
When measuring high resistance, a shielding box should be used to shield the object being measured.
15. ★ It is strictly prohibited to change the voltage range arbitrarily during the testing process and to turn on the host during the power on process.
16. ★ It is strictly prohibited to replace the sample if the current resistance range is not in the 104 range and the voltage is in the 10V range.

Operation method
The BEST-121 volume and surface resistivity tester is a precision instrument capable of measuring * resistance and very weak current because the machine has an ultra-high performance electrostatic amplifier inside. Despite various protective measures, this ultra-high performance electrostatic amplifier may be damaged or its measurement accuracy and performance may be reduced due to excessive current surge, overvoltage, or discharge during incorrect operation and use.
Have you read the consequences of incorrect operation during the testing process
Before power-on
Not set the voltage to 10V and the resistance current range to 104
Overcurrent impact causes damage to the instrument amplifier due to overcurrent
test process
Adjust the voltage range of the rear panel arbitrarily to change the test voltage
Instrument damage caused by overcurrent or reverse discharge shock
During the testing process, the surface resistance of the shielding box was changed, and the surface electrical testing switch was damaged due to overcurrent or reverse discharge impact on the instrument caused by overcurrent
During the testing process, the test voltage is too high, causing the test material to pass through the current or reverse discharge impact, resulting in instrument overcurrent and damage or reduced measurement performance
During the testing process, the microcurrent measuring line and the high-voltage line are short circuited, causing overcurrent or reverse discharge impact on the instrument, resulting in damage or reduced measurement performance
Test completed
The voltage range was not set to 10V and the resistance current range was not set to 104 for shutdown.
Overcurrent or reverse discharge impact causes damage to the instrument due to overcurrent or reduced measurement performance
Overview
This instrument can measure both ultra-high resistance and extremely weak current. By adopting large-scale integrated circuits and specialized technology, the instrument is compact, lightweight, and highly accurate. Display the resistor on a digital LCD and directly show the current flowing through the measured resistor. The resistance range is from 1 × 104 Ω to 1 × 1018 Ω, and the current measurement range is from 2 × 10-4A to 1 × 10-16A. The internal test voltages are DC10V, 50V, 100V, 250V, 500V, and 1000V.
This instrument has high accuracy, fast display, good stability, and convenient reading. It is suitable for testing the resistance values of anti-static products such as anti-static shoes, anti-static plastic and rubber products, anti-static raised floors in computer rooms, as well as measuring the insulation resistance of insulation materials and electronic and electrical products. This instrument can not only measure resistance, but also directly measure current such as dark current in electronic devices.
Two main characteristics
Resistance measurement range: 1 × 104 Ω~1 × 1018 Ω;
Current measurement range 2 × 10-4A to 1 × 10-16A;
Small size, light weight, and high accuracy;
The dual display of the measured resistance and the current flowing through the resistance makes operation and measurement more convenient;
Stable performance and convenient reading;
Can measure both resistance and current;
There are six voltage options for testing: DC10V, 50V, 100V, 250V, 500V, and 1000V;
Easy to use and operate, it can directly read and display digital results at any resistance range and test voltage, eliminating the trouble of multiplying by a coefficient, making measuring ultra-high resistance as simple as measuring ordinary resistance with a multimeter.
Insulation resistance tester
Semiconductor resistance tester
Vulcanized rubber insulation resistance tester
Vulcanized rubber volume surface insulation resistance tester
Vulcanized rubber resistivity tester
Plastic volume surface resistivity tester
Plastic insulation resistance tester
Plastic resistivity tester
Plastic volume resistivity tester
ZC36 Insulation Resistance Tester
Zc36 high resistance meter
ZC90 Insulation Resistance Tester
Volume surface resistivity tester
Volume resistivity tester
Surface resistivity tester
Direct resistivity tester
DC insulation resistance tester
Rubber insulation resistance tester
Insulation material resistance tester
Conductive material resistivity tester
High resistance measuring instrument
Powder resistivity tester
Liquid resistivity tester
Coating volume surface resistivity tester
Liquid plasticizer volume resistivity tester
GB1410 resistivity tester
Insulation paint resistivity tester
Thin film resistivity tester
Insulation paper resistivity tester
Plastic resistivity tester
Rubber resistivity tester
Solid insulation material volume surface resistivity tester
Vulcanized rubber insulation resistivity tester
Vulcanized rubber conductivity and dissipation performance resistivity tester
Oil tank static conductive coating resistivity tester
Coating volume surface resistivity tester
Insulation material high temperature resistivity tester
Textile resistivity tester
Liquid insulation material resistivity measuring instrument
Insulation paint volume surface resistivity tester
Petroleum wax and petroleum grease volume resistivity tester