Welcome Customer !

Membership

Help

Beijing AVIC Times Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Video
Video

Beijing AVIC Times Instrument Equipment Co., Ltd

  • E-mail

    zhsdyq@163.com

  • Phone

    13699145010

  • Address

    No.1 Economic and Technological Development Zone, Fangshan District, Beijing

Contact Now
Brief Introduction

ZJC series microcomputer controlled breakdown voltage tester product introduction (20KV-150KV)

1purpose

Mainly suitable for testing the breakdown strength and withstand time of solid insulation materials such as resins and adhesives, impregnated fiber products, mica and its products, plastic composite products, ceramics and glass under power frequency voltage or DC voltage; This instrument is computer-controlled and can quickly and accurately collect, process, access, display, and print various data during the experimental process.
2. According to the standard
The instrument is manufactured according to the requirements for breakdown voltage in the following standards
1. GB1408.1-2006 "Test Methods for Electrical Strength of Insulation Materials"
2. GB1408.2-2006 "Test Methods for Electrical Strength of Insulation Materials Part 2: Additional Requirements for Application of DC Voltage Test"
3. GB/T1695-2005 "Determination method for power frequency breakdown voltage strength and withstand voltage of vulcanized rubber"
4. ASTM D149 "Test Method for Dielectric Breakdown Voltage and Dielectric Strength of Solid Electrical Insulation Materials at Industrial Power Supply Frequency"

5. GB/T3333: Test method for power frequency voltage breakdown of cable paper;

6. GB/T12656: Test method for power frequency voltage breakdown of capacitor paper

7. HG/T3330: Determination method for breakdown strength of insulation paint film.

3. Function:

1. During the experiment, the experimental curve can be dynamically drawn, and the experimental curve can be overlaid and compared with multiple colors.
2. Can edit and modify experimental data, flexible and applicable;
3. The experimental conditions and test results can be automatically stored;
4. The format of the test report is flexible and adaptable to meet the different needs of different users;
5. The validity of curve data in a set of experiments can be manually selected;
6. The experimental result data can be imported into Excel and edited in WORD documents;
7. The overcurrent protection device has sufficient sensitivity to ensure that the power is cut off within 0.1 seconds when the sample breaks down;
8. Durability of instrument operation: The instrument can operate continuously without the need for periodic shutdowns to protect it. .

3、 Main technical requirements: (Parameters can be modified or customized according to usage requirements)

Project

ZJC-20KV

ZJC-50KV

ZJC-100KV

ZJC-150KV

input voltage

220V 50HZ

380V 50HZ

Voltage measurement range

AC/DC 0-20KV

AC/DC 0-50KV

AC/DC 0-100KV

AC/DC 0-150KV

Electrical capacity (power)

2KVA

5KVA

10KVA

15KVA

overcurrent protection

0-50mA

0-150mA

pressure rate

0.1KV/S-5KV/S

0.5KV/S-10KV/S

Experimental method

AC/DC test: 1. Constant voltage boost 2. Step by step boost 3. Voltage withstand test

Measurement error of AC/DC voltage

2%≤

Current measurement error

2%≤

Pressure resistance time

0-4H (German standard 8H or 15H) or any setting

Instrument dimensions (length, width, height)

700*800*1400mm

1200*1100*1500mm

2100*1500*2100mm

Host rescheduling

150kg

300KG

500kg

Level 9 security protection

Overvoltage, test overcurrent, test short circuit, safety door opening, software misoperation, zero voltage reset, discharge at the end of the test, independent protective grounding, electromagnetic discharge after the test is completed